Paced
AV sequential (dual-chamber) pacing
A dual-chamber pacemaker paces the atrium, waits a set AV delay, then paces the ventricle, preserving the normal atrial-then-ventricular sequence.
Last updated

Example tracing, drawn by the lab's engine. Open the lab for a live, adjustable version.
At a glance
- Rate
- The programmed rate, perfectly regular
- Pacing spikes
- Two per beat: atrial, then ventricular
- P waves
- Follow the first spike
- AV delay
- Fixed, usually 120–200 ms between spikes
Figures
Still images drawn by the lab's engine at 25 mm/s and 10 mm/mV, with the finding labeled. Select a figure to enlarge it.
What the heart is doing
The device stimulates the right atrium, times an AV delay that stands in for the AV node, then stimulates the right ventricle. Ventricular activation spreads through muscle, widening the QRS.
Why it looks this way
This is AV sequential pacing because each beat has two spikes: one before the P wave and, after a fixed delay, one before a wide QRS.
Why it matters
Normal dual-chamber function. Keeping the atrial contribution to filling improves cardiac output compared with ventricular pacing alone. Assess capture of both chambers.
What to look for
- Two spikes in each cycle, a constant distance apart.
- A P wave after the first spike and a wide QRS after the second.
- Check that both spikes capture on every beat.
Criteria
- Rate
- The programmed rate, perfectly regular
- Pacing spikes
- Two per beat: atrial, then ventricular
- P waves
- Follow the first spike
- AV delay
- Fixed, usually 120–200 ms between spikes
- QRS
- Wide, LBBB-like, after the second spike
Causes
- AV block with sinus node dysfunction
- Complete heart block in a patient in sinus rhythm
- Symptomatic bradycardia needing AV synchrony
Can be mistaken for
- Ventricular pacing alone — one spike per beat and no paced P wave Compare
- Atrial pacing — one spike per beat followed by a narrow QRS Compare
- Sinus rhythm with LBBB and artifact — no consistent spikes
Advanced Down to the ion channel
What the cells are doing
Pacemaker, atrial and ventricular action potentials, in step with the tracing above, with sodium, calcium and potassium moving through their channels.
Frequently asked questions
What are the ECG criteria for AV sequential (dual-chamber) pacing?
Rate: the programmed rate, perfectly regular. Pacing spikes: two per beat: atrial, then ventricular. P waves: follow the first spike. AV delay: fixed, usually 120–200 ms between spikes. QRS: wide, LBBB-like, after the second spike.
How is AV sequential (dual-chamber) pacing identified on an ECG?
This is AV sequential pacing because each beat has two spikes: one before the P wave and, after a fixed delay, one before a wide QRS.
What can AV sequential (dual-chamber) pacing be mistaken for?
Ventricular pacing alone: one spike per beat and no paced P wave. Atrial pacing: one spike per beat followed by a narrow QRS. Sinus rhythm with LBBB and artifact: no consistent spikes.
What causes AV sequential (dual-chamber) pacing?
Common causes: AV block with sinus node dysfunction; complete heart block in a patient in sinus rhythm; and symptomatic bradycardia needing AV synchrony.
Sources
This page does not list page-specific sources yet.
Every source the site cites, with links