Position & lead placement

Arm lead reversal (LA/RA)

The left-arm and right-arm electrodes have been swapped. Lead I is flipped upside down and aVR and aVL trade places, while the chest leads are unaffected.

Last updated

Live tracing25 mm/s · 10 mm/mV
Example ECG tracing of arm lead reversal (LA/RA) (simulated).

Example tracing, drawn by the lab's engine. Open the lab for a live, adjustable version.

At a glance

Lead I
P wave, QRS and T wave all inverted
aVR and aVL
Swapped: aVR looks upright
II and III
Swapped
aVF
Unchanged

All 5 criteria

Figures

Still images drawn by the lab's engine at 25 mm/s and 10 mm/mV, with the finding labeled. Select a figure to enlarge it.

Annotated, simulated ECG of arm lead reversal (LA/RA). Leads I, aVR and V6, the same beat, zoomed, with the finding labeled. Labels: Inverted P; Negative QRS; Upright aVR; Chest leads normal.
Annotated still. Leads I, aVR and V6, the same beat, zoomed, with the finding labeled.
Simulated 12-lead ECG of arm lead reversal (LA/RA), with a lead II rhythm strip. Tinted violet, where the finding shows: I, aVR, aVL.

Scroll sideways to see the whole tracing, or select it to enlarge.

12-lead. A full 12-lead with a lead II rhythm strip. Violet tint: where the finding shows in I, aVR and aVL.

What the heart is doing

Lead I measures left arm minus right arm, so swapping the electrodes reverses its polarity. The leads built from each arm exchange identities, while the chest electrodes are untouched.

Why it looks this way

This is arm lead reversal because lead I is completely inverted and aVR is upright while the precordial leads show normal R-wave progression.

Why it matters

The most common recording error. It carries no risk to the patient except misdiagnosis: it imitates lateral infarction, an ectopic atrial rhythm or dextrocardia. Always repeat the recording.

What to look for

  • An upright P wave and QRS in aVR is almost always a lead error.
  • Lead I and V6 normally look alike; if I is inverted and V6 is normal, the arm leads are swapped.
  • Check the electrodes and repeat the ECG before interpreting anything else.

Criteria

Lead I
P wave, QRS and T wave all inverted
aVR and aVL
Swapped: aVR looks upright
II and III
Swapped
aVF
Unchanged
V1–V6
Normal R-wave progression

Causes

  • Electrodes attached to the wrong arms
  • Cables plugged into the wrong electrodes
  • Tangled or unlabeled lead wires
  • Hurried acquisition in an emergency

Can be mistaken for

  • Dextrocardia — the R wave also shrinks from V1 to V6 Compare
  • Ectopic atrial rhythm — an abnormal P axis can invert P in lead I, but QRS and T stay upright Compare
  • High lateral infarction — Q waves in I and aVL with an upright P wave

Advanced Down to the ion channel

What the cells are doing

Pacemaker, atrial and ventricular action potentials, in step with the tracing above, with sodium, calcium and potassium moving through their channels.

Frequently asked questions

What are the ECG criteria for arm lead reversal (LA/RA)?

Lead I: P wave, QRS and T wave all inverted. aVR and aVL: swapped: aVR looks upright. II and III: swapped. aVF: unchanged. V1–V6: normal R-wave progression.

How is arm lead reversal (LA/RA) identified on an ECG?

This is arm lead reversal because lead I is completely inverted and aVR is upright while the precordial leads show normal R-wave progression.

What can arm lead reversal (LA/RA) be mistaken for?

Dextrocardia: the R wave also shrinks from V1 to V6. Ectopic atrial rhythm: an abnormal P axis can invert P in lead I, but QRS and T stay upright. High lateral infarction: Q waves in I and aVL with an upright P wave.

What causes arm lead reversal (LA/RA)?

Common causes: electrodes attached to the wrong arms; cables plugged into the wrong electrodes; tangled or unlabeled lead wires; and hurried acquisition in an emergency.