Electrical therapy: cardioversion, defibrillation and pacing
Defibrillation is an unsynchronized shock for rhythms with no organized QRS to time against. Synchronized cardioversion times the shock to the QRS so it does not land on the T wave. Transcutaneous pacing supports a symptomatic bradycardia until the cause is treated or a transvenous wire is placed.
Guidelines: AHA 2020 with the 2023 update and confirmed 2025 changes; PALS 2020 with the 2025 pediatric updates
Reflects the AHA 2020 adult algorithms with the 2023 focused update, updated for the 2025 AHA Guidelines for CPR and ECC where the change could be confirmed; acute coronary syndromes follow the 2025 ACC/AHA guideline and stroke the 2026 AHA/ASA guideline. Pediatric content follows the AHA 2020 PALS algorithms with the 2025 AHA/AAP pediatric updates where confirmed.
Walk through it
One step at a time. Each step says what to do and why; at a decision, choose the branch that fits your patient. The map shows where you are.
The interactive walk needs JavaScript. Every step is listed below.
Every step at once
Start
What do you need the electricity to do?
Why: Defibrillation, cardioversion and pacing use the same pads but answer three different problems: no pulse, too fast with a pulse, too slow with a pulse.
Then
Defibrillation (unsynchronized)
- Biphasic: manufacturer's energy (typically 120–200 J), or the maximum if unknown. Monophasic: 360 J.
- Compressions continue while charging; clear everyone; shock; resume compressions at once.
- Pads: anterior–lateral or anterior–posterior.
Why: A shock depolarizes a critical mass of myocardium at once. With every cell refractory together, the chaotic wavelets have nowhere to go, and the sinus node can take over.
Rhythm: Ventricular fibrillation, coarse
Decide
Is there an organized QRS to synchronize to?
Why: Synchronization needs a clear R wave on every beat. Without one, the device may never fire.
Then
Polymorphic VT: defibrillate
- Treat as VF if unstable or pulseless: unsynchronized, defibrillation energy.
Why: The QRS changes shape and size beat to beat, so the sync marker cannot reliably find the R wave, and a delayed shock costs time.
Rhythm: Torsades de pointes
Do
Synchronized cardioversion
- Press SYNC and confirm a marker on every R wave.
- Sedate if time allows.
- AF and flutter: at least 200 J biphasic (2025). Narrow regular and monomorphic VT: follow the device (commonly 100 J).
- Hold the shock button: the device waits for the next R wave.
Why: A shock that lands on the T wave, during the vulnerable period of repolarization, can itself cause VF. Timing it to the R wave avoids that.
Then
If another shock is needed
- Many devices return to unsynchronized after each shock: press SYNC again.
- Increase the energy stepwise.
- If the rhythm degenerates into VF: turn sync off and defibrillate.
Why: Forgetting to re-sync is the classic error; forgetting to turn it off in VF means the device will not fire at all.
Do
Transcutaneous pacing
- Pads anterior–posterior.
- Set the rate (commonly 60–80/min).
- Increase the output (mA) from low until every spike is followed by a wide QRS, then set it a little above that threshold.
Why: The current has to cross the chest wall to reach the heart. Too little and the spikes fall on nothing; at threshold every spike captures.
Rhythm: Pacemaker failure to capture
Then: Electrical capture: is every spike followed by a wide QRS and a broad T wave?
Decide
Electrical capture: is every spike followed by a wide QRS and a broad T wave?
Why: A spike alone proves nothing: large pacing spikes can hide or imitate a QRS. Look for the T wave that follows a real paced beat.
Do
No capture: raise the output
- Increase mA; check pad contact and position; try anterior–posterior.
Why: Thick chest walls, air, effusion and poor pad contact all raise the threshold.
Then: Electrical capture: is every spike followed by a wide QRS and a broad T wave?
Decide
Mechanical capture: is there a femoral pulse matching the paced rate?
Why: Electrical capture without a pulse is PEA. Carotid pulses are unreliable here: muscle twitch from the pads can feel like one.
Then
Capture confirmed
- Analgesia and sedation.
- Recheck capture often; arrange transvenous pacing.
Why: Capture can be lost as the patient moves or sweats.
Rhythm: Ventricular paced rhythm
Then
Electrical capture without a pulse
- This is PEA: start CPR and follow the cardiac arrest algorithm.
- Transcutaneous pacing is not used for asystole in arrest.
Why: The paced complexes are electrical only; the heart is not pumping.
The rhythms
Paced with capture
A spike followed every time by a wide QRS and a T wave.
Entry: Ventricular paced rhythm
Pacing without consistent capture
Spikes with no QRS after some of them.
Entry: Pacemaker failure to capture
At a glance
The whole algorithm on one page. Read from the top; a decision splits the path.
Decide
Is there an organized QRS to synchronize to?
No: VF, pulseless VT, polymorphic VT
Do
Defibrillate (unsynchronized)
- Polymorphic VT is too irregular to synchronize reliably: treat it as VF if the patient is unstable.
Yes: unstable SVT, AF, flutter, monomorphic VT with a pulse
Do
Synchronized cardioversion
- Press sync; confirm a marker on every QRS.
- Sedate if possible.
- Many devices reset to unsynchronized after each shock: re-enable sync before the next one.
Decide
Symptomatic bradycardia not responding to atropine?
Yes
Do
Transcutaneous pacing
- Pads front and back (or front and side).
- Set the rate and increase the output until every spike captures, then set it slightly above that threshold as the device and protocol direct.
- Check electrical capture on the monitor and mechanical capture by pulse.
No
Then
Continue the bradycardia algorithm
- Treat the cause; observe; consider expert consultation.
Key points
Electrical capture
- Every pacing spike is followed by a wide QRS complex.
- The paced QRS has a broad T wave pointing the opposite way.
- Large transcutaneous spikes can hide or mimic a QRS: look for the T wave.
Mechanical capture
- A pulse that matches the paced rate: feel the femoral pulse, as muscle twitch from the pads can be mistaken for a carotid pulse.
- A pulse oximetry or arterial waveform at the paced rate, and better perfusion.
Pacing indications (with a pulse)
- Symptomatic bradycardia when atropine is ineffective or not appropriate.
- High-grade AV block (Mobitz II or third degree) with poor perfusion: do not delay pacing for atropine.
- Not for asystole during cardiac arrest: pacing does not improve outcome there.
Energies, doses and targets
| What | As the guideline states it | Source |
|---|---|---|
| Defibrillation, biphasic | Manufacturer recommendation (for example, initial dose of 120–200 J); if unknown, use the maximum available. Second and later doses should be equivalent; higher doses may be considered. | AHA 2020 adult algorithm |
| Defibrillation, monophasic | 360 J | AHA 2020 adult algorithm |
| Synchronized cardioversion | Refer to your specific device's recommended energy level to maximize first-shock success. | AHA 2020 adult algorithm |
| Synchronized cardioversion of AF (2025 change) | Initial energy of at least 200 J with a biphasic defibrillator | AHA 2025 Guidelines, Part 9: Adult Advanced Life Support (as reported in the 2025 highlights) |
Mechanisms, cautions and pediatric doses: the drugs. Practise it: megacode cases.
Guidelines and sources
Reflects the AHA 2020 adult algorithms with the 2023 focused update, updated for the 2025 AHA Guidelines for CPR and ECC where the change could be confirmed; acute coronary syndromes follow the 2025 ACC/AHA guideline and stroke the 2026 AHA/ASA guideline.
Algorithms and doses follow the AHA 2020 Guidelines for CPR and ECC, adult algorithms (Part 3: Adult Basic and Advanced Life Support, Circulation 2020;142(suppl 2):S366–S468), checked for changes in:
- 2023 AHA Focused Update on Adult Advanced Cardiovascular Life Support (Circulation 2024;149:e254–e273)
- 2025 AHA Guidelines for CPR and ECC, Part 9: Adult Advanced Life Support
Changes in the 2025 guidelines reflected here:
- Synchronized cardioversion of atrial fibrillation: an initial energy of at least 200 J with a biphasic defibrillator is now recommended, rather than a lower first shock.
- The tachycardia algorithm refers to the device's recommended energy level rather than listing energies for each rhythm.
- Intravenous access is preferred to intraosseous access as the first attempt in cardiac arrest.
- Double sequential defibrillation and vector-change defibrillation for refractory VF remain of uncertain benefit.
- Sotalol is no longer listed in the tachycardia algorithm; procainamide and amiodarone remain.
- Routine calcium and sodium bicarbonate are not recommended in cardiac arrest; both are kept for specific causes.
- After ROSC: titrate oxygen to an SpO₂ of 90–98%, keep the MAP at 65 mm Hg or more, and use deliberate temperature control (32–37.5 °C) for patients not following commands.
- Polymorphic VT is treated as unstable and defibrillated.
Sources:
- Panchal AR, et al. Part 3: Adult Basic and Advanced Life Support: 2020 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2020;142(16 suppl 2):S366–S468.
- Perman SM, et al. 2023 American Heart Association Focused Update on Adult Advanced Cardiovascular Life Support. Circulation. 2024;149:e254–e273.
- Lavonas EJ, et al. 2023 American Heart Association Focused Update on the Management of Patients With Cardiac Arrest or Life-Threatening Toxicity Due to Poisoning. Circulation. 2023;148:e149–e184.
- Part 7: Adult Basic Life Support. 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2).
- Part 9: Adult Advanced Life Support. 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2):S538–S577.
- Part 10: Adult and Pediatric Special Circumstances of Resuscitation. 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2).
- Part 11: Post–Cardiac Arrest Care. 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2).
- Rao SV, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI Guideline for the Management of Patients With Acute Coronary Syndromes. Circulation. 2025;151(13):e771–e862.
- Prabhakaran S, et al. 2026 Guideline for the Early Management of Patients With Acute Ischemic Stroke (AHA/ASA). Stroke. 2026. doi:10.1161/STR.0000000000000513.
The full 2025 guideline text could not be opened when this page was drafted; the 2025 changes listed were confirmed from summaries of AHA's published 2025 algorithms and highlights. Drug doses are the 2020 algorithm doses, which those 2025 algorithms keep. All text and pictures here are original teaching material, not reproductions of AHA algorithms. Education only, not a treatment protocol.